Filtered by vendor Philips Subscriptions
Filtered by product Intellivue Mx100 Subscriptions
Total 5 CVE
CVE Vendors Products Updated CVSS v3.1
CVE-2020-16228 1 Philips 24 Intellivue Mp2-mp90, Intellivue Mp2-mp90 Firmware, Intellivue Mx100 and 21 more 2024-11-21 6.4 Medium
In Patient Information Center iX (PICiX) Versions C.02 and C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX550, MX750, MX850, and IntelliVue X3 Versions N and prior, the software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate.
CVE-2020-16216 1 Philips 24 Intellivue Mp2-mp90, Intellivue Mp2-mp90 Firmware, Intellivue Mx100 and 21 more 2024-11-21 6.5 Medium
In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly, which can induce a denial-of-service condition through a system restart.
CVE-2018-10601 1 Philips 36 Avalon Fetal\/maternal Monitors Fm20, Avalon Fetal\/maternal Monitors Fm20 Firmware, Avalon Fetal\/maternal Monitors Fm30 and 33 more 2024-11-21 8.2 High
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
CVE-2018-10599 1 Philips 36 Avalon Fetal\/maternal Monitors Fm20, Avalon Fetal\/maternal Monitors Fm20 Firmware, Avalon Fetal\/maternal Monitors Fm30 and 33 more 2024-11-21 N/A
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to read memory from an attacker-chosen device address within the same subnet.
CVE-2018-10597 1 Philips 36 Avalon Fetal\/maternal Monitors Fm20, Avalon Fetal\/maternal Monitors Fm20 Firmware, Avalon Fetal\/maternal Monitors Fm30 and 33 more 2024-11-21 8.3 High
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to access memory ("write-what-where") from an attacker-chosen device address within the same subnet.