Show plain JSON{"dataType": "CVE_RECORD", "dataVersion": "5.1", "cveMetadata": {"cveId": "CVE-2024-9834", "assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa", "state": "PUBLISHED", "assignerShortName": "Baxter", "dateReserved": "2024-10-10T19:24:48.834Z", "datePublished": "2024-11-14T20:57:22.734Z", "dateUpdated": "2024-11-15T15:10:40.157Z"}, "containers": {"cna": {"affected": [{"defaultStatus": "unaffected", "product": "Life2000 Ventilation System", "vendor": "Baxter", "versions": [{"status": "affected", "version": "06.08.00.00 and prior"}]}], "datePublic": "2024-11-14T20:43:00.000Z", "descriptions": [{"lang": "en", "supportingMedia": [{"base64": false, "type": "text/html", "value": "Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.<br>"}], "value": "Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance."}], "impacts": [{"capecId": "CAPEC-117", "descriptions": [{"lang": "en", "value": "CAPEC-117 Interception"}]}, {"capecId": "CAPEC-441", "descriptions": [{"lang": "en", "value": "CAPEC-441 Malicious Logic Insertion"}]}], "metrics": [{"cvssV3_1": {"attackComplexity": "LOW", "attackVector": "LOCAL", "availabilityImpact": "HIGH", "baseScore": 9.3, "baseSeverity": "CRITICAL", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "CHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H", "version": "3.1"}, "format": "CVSS", "scenarios": [{"lang": "en", "value": "GENERAL"}]}], "problemTypes": [{"descriptions": [{"cweId": "CWE-319", "description": "CWE-319 Cleartext Transmission of Sensitive Information", "lang": "en", "type": "CWE"}]}], "providerMetadata": {"orgId": "dba971b9-eb30-4121-91e1-3b45611354aa", "shortName": "Baxter", "dateUpdated": "2024-11-14T21:47:25.133Z"}, "references": [{"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"}], "source": {"discovery": "UNKNOWN"}, "title": "Improper data protection on Life2000 ventilator serial interface", "x_generator": {"engine": "Vulnogram 0.2.0"}}, "adp": [{"affected": [{"vendor": "baxter", "product": "life2000_ventilator_firmware", "cpes": ["cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"], "defaultStatus": "unaffected", "versions": [{"version": "0", "status": "affected", "lessThanOrEqual": "06.08.00.00", "versionType": "custom"}]}], "metrics": [{"other": {"type": "ssvc", "content": {"timestamp": "2024-11-15T15:09:40.431805Z", "id": "CVE-2024-9834", "options": [{"Exploitation": "none"}, {"Automatable": "no"}, {"Technical Impact": "total"}], "role": "CISA Coordinator", "version": "2.0.3"}}}], "title": "CISA ADP Vulnrichment", "providerMetadata": {"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP", "dateUpdated": "2024-11-15T15:10:40.157Z"}}]}}