Show plain JSON{"containers": {"cna": {"affected": [{"product": "Baxter Sigma Spectrum Infusion Pumps", "vendor": "n/a", "versions": [{"status": "affected", "version": "Sigma Spectrum v6.x model 35700BAX,Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module (WBM) v9,v11,v13,v14,v15,v16,v20D29,v20D30,v20D31,v22D24, Spectrum v8.x w/WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum LVP v8.x with WBM v17, v20D29,v20D30,v20D31,and v22D24"}]}], "descriptions": [{"lang": "en", "value": "The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials."}], "problemTypes": [{"descriptions": [{"cweId": "CWE-259", "description": "USE OF HARD-CODED PASSWORD CWE-259", "lang": "en", "type": "CWE"}]}], "providerMetadata": {"dateUpdated": "2020-06-29T13:41:37", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert"}, "references": [{"tags": ["x_refsource_MISC"], "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04"}], "x_legacyV4Record": {"CVE_data_meta": {"ASSIGNER": "ics-cert@hq.dhs.gov", "ID": "CVE-2020-12047", "STATE": "PUBLIC"}, "affects": {"vendor": {"vendor_data": [{"product": {"product_data": [{"product_name": "Baxter Sigma Spectrum Infusion Pumps", "version": {"version_data": [{"version_value": "Sigma Spectrum v6.x model 35700BAX,Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module (WBM) v9,v11,v13,v14,v15,v16,v20D29,v20D30,v20D31,v22D24, Spectrum v8.x w/WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum LVP v8.x with WBM v17, v20D29,v20D30,v20D31,and v22D24"}]}}]}, "vendor_name": "n/a"}]}}, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": {"description_data": [{"lang": "eng", "value": "The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials."}]}, "problemtype": {"problemtype_data": [{"description": [{"lang": "eng", "value": "USE OF HARD-CODED PASSWORD CWE-259"}]}]}, "references": {"reference_data": [{"name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04", "refsource": "MISC", "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04"}]}}}, "adp": [{"providerMetadata": {"orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE", "dateUpdated": "2024-08-04T11:48:57.938Z"}, "title": "CVE Program Container", "references": [{"tags": ["x_refsource_MISC", "x_transferred"], "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04"}]}]}, "cveMetadata": {"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2020-12047", "datePublished": "2020-06-29T13:41:37", "dateReserved": "2020-04-21T00:00:00", "dateUpdated": "2024-08-04T11:48:57.938Z", "state": "PUBLISHED"}, "dataType": "CVE_RECORD", "dataVersion": "5.1"}