Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
History

No history.

cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published: 2018-08-13T22:00:00Z

Updated: 2024-09-17T03:43:42.929Z

Reserved: 2018-08-01T00:00:00