Show plain JSON{"containers": {"cna": {"affected": [{"defaultStatus": "unaffected", "product": "8840 N\u2019Vision Clinician Programmer", "vendor": "Medtronic", "versions": [{"status": "affected", "version": "All versions"}]}, {"defaultStatus": "unaffected", "product": "8870 N\u2019Vision removable Application Card", "vendor": "Medtronic", "versions": [{"status": "affected", "version": "All versions"}]}], "credits": [{"lang": "en", "type": "finder", "value": "Billy Rios and Jonathan Butts of Whitescope LLC reported this vulnerability"}], "descriptions": [{"lang": "en", "supportingMedia": [{"base64": false, "type": "text/html", "value": "<p>\n\n<span style=\"background-color: rgb(255, 255, 255);\">\nMedtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programmer and 8870 N'Vision removable application card does not encrypt PII and PHI while at rest.\n\n</span>\n\n</p>"}], "value": "Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programmer and 8870 N'Vision removable application card does not encrypt PII and PHI while at rest."}], "metrics": [{"cvssV3_1": {"attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "NONE", "baseScore": 4.6, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N", "version": "3.1"}, "format": "CVSS", "scenarios": [{"lang": "en", "value": "GENERAL"}]}], "problemTypes": [{"descriptions": [{"cweId": "CWE-311", "description": "CWE-311 Missing Encryption of Sensitive Data", "lang": "en", "type": "CWE"}]}], "providerMetadata": {"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert", "dateUpdated": "2025-05-22T17:56:24.735Z"}, "references": [{"url": "https://global.medtronic.com/xg-en/product-security/security-bulletins/nvision.html"}, {"url": "https://www.medtronic.com/security"}, {"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01"}, {"url": "http://www.securityfocus.com/bid/104213"}], "source": {"advisory": "ICSMA-18-137-01", "discovery": "EXTERNAL"}, "title": "Medtronic N'Vision Clinician Programmer Missing Encryption of Sensitive Data", "workarounds": [{"lang": "en", "supportingMedia": [{"base64": false, "type": "text/html", "value": "<p>Medtronic has not developed a product update to address the vulnerabilities, but is reinforcing security reminders within this advisory to help reduce the risk associated with the vulnerabilities.</p><p>The 8870 Therapy Application card stores PHI and PII as part of its normal operating procedure and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy.</p><p>Medtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, hospitals and clinicians should:</p><ul><li>Maintain strict physical control of the 8870 application card.</li><li>Use only legitimately obtained 8870 cards and not cards provided by any third party as firmware and system updates are provided directly by Medtronic using new 8870 application cards.</li><li>8840 Programmers and 8870 Therapy Application compact flash cards are the property of Medtronic and should be returned to Medtronic when no longer in use. If that is not an option, you should securely dispose of them.</li></ul><p>Medtronic has released additional patient focused information, at the following location:</p><p><a target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\">https://www.medtronic.com/security</a></p>"}], "value": "Medtronic has not developed a product update to address the vulnerabilities, but is reinforcing security reminders within this advisory to help reduce the risk associated with the vulnerabilities.\n\nThe 8870 Therapy Application card stores PHI and PII as part of its normal operating procedure and should be handled, managed and secured in a manner consistent with the applicable laws for patient data privacy.\n\nMedtronic recommends users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, hospitals and clinicians should:\n\n * Maintain strict physical control of the 8870 application card.\n * Use only legitimately obtained 8870 cards and not cards provided by any third party as firmware and system updates are provided directly by Medtronic using new 8870 application cards.\n * 8840 Programmers and 8870 Therapy Application compact flash cards are the property of Medtronic and should be returned to Medtronic when no longer in use. If that is not an option, you should securely dispose of them.\n\n\nMedtronic has released additional patient focused information, at the following location:\n\n https://www.medtronic.com/security"}], "x_ConverterErrors": {"DATE_PUBLIC": {"error": "v4 DATE_PUBLIC is invalid", "message": "month must be in 1..12"}}, "x_generator": {"engine": "Vulnogram 0.2.0"}, "x_legacyV4Record": {"CVE_data_meta": {"ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2017-50-17T00:00:00", "ID": "CVE-2018-10631", "STATE": "PUBLIC"}, "affects": {"vendor": {"vendor_data": [{"product": {"product_data": [{"product_name": "Medtronic N'Vision Clinician Programmer", "version": {"version_data": [{"version_value": "8840 N'Vision Clinician Programmer, all versions, and 8870 N'Vision removable Application Card, all versions."}]}}]}, "vendor_name": "ICS-CERT"}]}}, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": {"description_data": [{"lang": "eng", "value": "Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programmer, all versions, and 8870 N'Vision removable Application Card, all versions. The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card, including the binary executables. If modified to bypass protection mechanisms, this malicious code will be run when the card is inserted into an 8840 Clinician Programmer."}]}, "problemtype": {"problemtype_data": [{"description": [{"lang": "eng", "value": "PROTECTION MECHANISM FAILURE CWE-693"}]}]}, "references": {"reference_data": [{"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01"}, {"name": "https://www.medtronic.com/security", "refsource": "MISC", "url": "https://www.medtronic.com/security"}]}}}, "adp": [{"providerMetadata": {"orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE", "dateUpdated": "2024-08-05T07:46:46.502Z"}, "title": "CVE Program Container", "references": [{"tags": ["x_refsource_MISC", "x_transferred"], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01"}, {"tags": ["x_refsource_MISC", "x_transferred"], "url": "https://www.medtronic.com/security"}]}, {"problemTypes": [{"descriptions": [{"type": "CWE", "cweId": "CWE-284", "lang": "en", "description": "CWE-284 Improper Access Control"}]}], "affected": [{"vendor": "ics_cert", "product": "medtronic_n_vision_clinician_programmer", "cpes": ["cpe:2.3:a:ics_cert:medtronic_n_vision_clinician_programmer:*:*:*:*:*:*:*:*"], "defaultStatus": "unknown", "versions": [{"version": "8840_n_vision_clinician_programmer", "status": "affected", "lessThanOrEqual": "*", "versionType": "custom"}, {"version": "8870_n_vision_removable_application_card", "status": "affected"}]}], "metrics": [{"cvssV3_1": {"scope": "UNCHANGED", "version": "3.1", "baseScore": 6.6, "attackVector": "PHYSICAL", "baseSeverity": "MEDIUM", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H", "integrityImpact": "HIGH", "userInteraction": "REQUIRED", "attackComplexity": "LOW", "availabilityImpact": "HIGH", "privilegesRequired": "NONE", "confidentialityImpact": "HIGH"}}, {"other": {"type": "ssvc", "content": {"timestamp": "2024-08-19T18:07:51.945413Z", "id": "CVE-2018-10631", "options": [{"Exploitation": "none"}, {"Automatable": "no"}, {"Technical Impact": "total"}], "role": "CISA Coordinator", "version": "2.0.3"}}}], "title": "CISA ADP Vulnrichment", "providerMetadata": {"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0", "shortName": "CISA-ADP", "dateUpdated": "2024-08-19T18:26:09.940Z"}}]}, "cveMetadata": {"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2018-10631", "dateReserved": "2018-05-01T00:00:00", "dateUpdated": "2025-05-22T17:56:24.735Z", "state": "PUBLISHED", "datePublished": "2018-07-13T19:00:00Z"}, "dataType": "CVE_RECORD", "dataVersion": "5.1"}