Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
History

No history.

cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published: 2019-03-25T16:12:01

Updated: 2024-08-06T06:04:01.128Z

Reserved: 2015-05-12T00:00:00

Link: CVE-2015-3954

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Modified

Published: 2019-03-25T17:29:00.670

Modified: 2024-11-21T02:30:08.203

Link: CVE-2015-3954

cve-icon Redhat

No data.