Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Metrics
Affected Vendors & Products
References
Link | Providers |
---|---|
https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01 |
History
No history.
MITRE
Status: PUBLISHED
Assigner: icscert
Published: 2019-03-25T16:12:01
Updated: 2024-08-06T06:04:01.128Z
Reserved: 2015-05-12T00:00:00
Link: CVE-2015-3954
Vulnrichment
No data.
NVD
Status : Modified
Published: 2019-03-25T17:29:00.670
Modified: 2024-11-21T02:30:08.203
Link: CVE-2015-3954
Redhat
No data.