Filtered by vendor Bbraun
Subscriptions
Filtered by product Spacestation 8713142u
Subscriptions
Total
5 CVE
CVE | Vendors | Products | Updated | CVSS v3.1 |
---|---|---|---|---|
CVE-2021-33886 | 1 Bbraun | 3 Infusomat Large Volume Pump 871305u, Spacecom2, Spacestation 8713142u | 2024-11-21 | 8.1 High |
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device. | ||||
CVE-2021-33885 | 1 Bbraun | 3 Infusomat Large Volume Pump 871305u, Spacecom2, Spacestation 8713142u | 2024-11-21 | 10 Critical |
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets. | ||||
CVE-2021-33884 | 1 Bbraun | 3 Infusomat Large Volume Pump 871305u, Spacecom2, Spacestation 8713142u | 2024-11-21 | 6.5 Medium |
An Unrestricted Upload of File with Dangerous Type vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows remote attackers to upload any files to the /tmp directory of the device through the webpage API. This can result in critical files being overwritten. | ||||
CVE-2021-33883 | 1 Bbraun | 3 Infusomat Large Volume Pump 871305u, Spacecom2, Spacestation 8713142u | 2024-11-21 | 5.9 Medium |
A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to obtain sensitive information by snooping on the network traffic. The exposed data includes critical values for a pump's internal configuration. | ||||
CVE-2021-33882 | 1 Bbraun | 3 Infusomat Large Volume Pump 871305u, Spacecom2, Spacestation 8713142u | 2024-11-21 | 6.8 Medium |
A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands. |
Page 1 of 1.